Specifications:
Measurement Function Range Frequency Range/ Testing Current Accuracy:1 year ± ( % of reading + % of range)
DC Voltage 200mV ∼ 1000V   0.015 + 0.003
DC Current 200uA ∼ 10A   0.020 + 0.005
AC Voltage(RMS) 200mV ∼ 750V 20Hz ∼ 100kHz 0.20 + 0.05
AC Current(RMS) 20mA ∼ 10A 20Hz ∼ 10kHz 0.30 + 0.06
Resistance
(2-wire and 4-wire)
200Ω ∼ 100MΩ   0.020 + 0.003
Capacitance 2nF ∼ 10000uF   1.0 + 0.5
Diode test 2.4V 1mA 0.05 + 0.01
  Frequency and Period 200mV ∼ 750V 10Hz ∼ 1MHz 0.01 + 0.003
20mA ∼ 10A 10Hz ∼ 100kHz 0.01 + 0.003
Continuity 2KΩ 1mA 0.05 + 0.01
note:All the indicators are the typical value under standard test situation
Other parameters
Measurement Function DC Voltage、DC Current、AC Voltage(RMS)、AC Current(RMS)、Resistance(2-wire and 4-wire)、Capacitance、Diodes test、Frequency and Period、Continuity、short current、any sensor
Math “Pass/Fail” Limit Test、Standard Deviation、Histogram、Relatively、 Null, Max/Min/Avg, dBm, dB
Other Functions Built-in 10 groups of configuration storage,10 groups of configuration storage of any sensor, 2048 historical reading data record and check, 10 groups of historical datum storage, Exterior trigger input and VMC output,Reading hold,Single trigger
Display Characteristic Multi-display, menu, multi-language help and waveform display
Security CAT II 600V,CAT I 1000V. Contaminated Class 2
Shock and Vibration MIL-T-28800, type III,class 5(sine)
Power Supply 110/220V ±20%, 45 ∼ 65Hz, 20W Max